Electrical Delay - HP 8712ET User Manual

Rf network analyzers
Hide thumbs Also See for HP 8712ET:
Table of Contents

Advertisement

TIP
NOTE
Optimizing Measurements
Compensating for Phase Shift in Measurement Setups
To use the port extension feature, press
Port Extensions Port Ext's on OFF
turned on, you can add delay independently to both the reflection port
and the transmission port.
To add delay between the REFLECTION RF OUT port and your DUT,
press
Refl Port Extension
To add delay between the TRANSMISSION RF IN port and your DUT,
press
Trans Port Extension
A good way to know when you have added enough delay is to connect a
short in place of the DUT. Adjust the port extension until the phase
response is flat. (Remember that most short-circuit standards have a
nonzero delay, so adjusting delay using this method will result in a delay
error equal to two times the short delay.)
The delay values that you enter will be automatically applied
appropriately to both transmission and reflection measurements. When
measuring reflection, the reflection port extension delay is applied twice.
When measuring transmission, the reflection port extension delay and
the transmission port extension delay are each applied once.

Electrical Delay

Another type of reference plane extension is
accessible from the
your current measurement to compensate for phase shift. To flatten the
phase response at a certain frequency, use
Marker Functions
adjusts the electrical delay to compensate for the phase slope at the
active marker position.
Electrical delay is commonly used in transmission measurements to
measure a device's electrical length, or to remove phase slope to measure
deviation from linear phase.
Unlike port extension, electrical delay does not automatically adjust the
applied delay when you switch between transmission and reflection
measurements. For this reason, port extensions are preferred over
electrical delay for fixture compensation or when measuring one-port
devices.
Reference plane extension only affects measurements using the
narrowband detectors.
5-18
, and enter the delay value.
, and enter the delay value.
key. Electrical delay lets you add delay to
SCALE
Marker -> Elec Delay
More Cal
,
CAL
. When port extension is
Electrical Delay
MARKER
. This automatically
ET User's Guide
,

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8714et

Table of Contents