HP 8712ET User Manual page 398

Rf network analyzers
Hide thumbs Also See for HP 8712ET:
Table of Contents

Advertisement

Table 9-1
System Performance, Corrected, 50Ω, T/R Calibration
Directivity
Source Match:
Reflection (One-Port Cal)
Transmission (Enhanced Response Cal)
Transmission (Response Cal)
Load Match
Reflection Tracking
Transmission Tracking:
Enhanced Response Cal
Response Cal
a. Option 1E1 adds a 60 dB step attenuator to the HP 8712ET/8714ET.
b. These specifications apply for measurements made using "fine" bandwidth,
Specifications
HP 8712ET/8714ET (50 Ω) with Attenuator Option 1E1
HP 85032B/E (Type-N, 50 Ω) Cal Kit, T/R Calibration
Description
no averaging, and at an environmental temperature of 25° ± 5° C, with less
than 1° C deviation from the calibration temperature.
9-4
a
b
Specification
(in dB)
300 kHz
1.3 GHz
to 1.3 GHz
to 3 GHz
50
47
42
36
42
36
21
15
18
15
±
±
0.02
0.02
±
±
0.039
0.055
±
±
0.13
0.3