Learning To Make Reflection Measurements - HP 8753E Installation And Quick Start Manual

Network analyzer
Hide thumbs Also See for HP 8753E:
Table of Contents

Advertisement

Learning to Make Reflection Measurements
This example procedure shows you how to measure the reflection response of a bandpass
filter. The measurement parameter values listed are unique to this particular test device.
For further measurement examples, refer to the "Making Measurements" chapter in the
HP 8753E Network Analyzer User's Guide.
NOTE
Reflection measurements monitor only one port of a test device. When a
test device has more than one port, you must ensure that the unused
port(s) are terminated in their characteristic impedance (for example,
50Ω or 75Ω). If you do not terminate unused ports, reflections from
these ports will cause measurement errors.
You can connect an unused device port to the unused analyzer test port,
to act as a termination. For example: when you are measuring S
S
22
PORT 1, respectively.
The signal reflected from the device under test is measured as a ratio of the reflected
energy versus the incident energy. It can be expressed as reflection coefficient, return loss,
or standing-wave-ratio (SWR). These measurements are mathematically defined as
follows:
reflection coefficient (Γ)
magnitude of reflection
coefficient (ρ)
return loss (dB)
standing-wave-ratio (SWR)
Chapter 2
, you can connect an unused device port to the analyzer's PORT 2 or
= reflected voltage / incident voltage
= S
or S
11
= |Γ|
= −20 log (ρ), where (ρ) = |Γ|
= V
maximum
= (1 + ρ) / (1 − ρ)
Quick Start: Learning How to Make Measurements
Learning to Make Reflection Measurements
(magnitude and phase)
22
/ V
minimum
or
11
41

Advertisement

Table of Contents
loading

Table of Contents