Optics Cleaning - Honeywell Vertex M Technical Handbook

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Vertex M
24-Point Continuous Monitor
TM
Note:
Some backup media may change the files to "read
only" and may require an additional step of chang-
ing the "Attributes" within the file properties to make
them usable.
Empty databases for the Event, Concentration and
ChemCam databases can be found in the Databases
folder on the Vertex M Technical Manual CD.
File Name
*.za_Vt
Configuration Profiles
ODBC Datasource for Event
CM72Data.dsn
Database
ODBC Datasource for
ConcData.dsn
Concentration Database
ODBC Datasource for
ChemCam.dsn
ChemCam Database
CM72.mdb *
Event database
CM72Conc.
Concentration Database
mbd*
ChemCam.
ChemCam Database
mdb *
* Required if stored locally on the Vertex M System
NOTE: Database files are size limited to a maximum of 2Gb.
Table 5-2. File Locations

5.11 Optics Cleaning

Clean Chemcassette
optics verification error occurs.
Required equipment:
Compressed air
Note:
Do not use "canned air" near Pyrolyzer gas based
detectors, the halogenated gasses may trigger
alarms. Instead, use N
is free of oils and contaminants.
Open the Optics Block Gate
Section 5 - Maintenance
Purpose
Location
C:\HMI\P_Util
C:\HMI
C:\HMI
C:\HMI
C:\HMI
C:\HMI
C:\HMI\
ChemCam
optics annually or whenever
®
or a compressed source that
2
Remove the Chemcassette
Open the Vertex M side panel
Remove tubing (shown in photo) one at the
time and blow out with compressed air.
Re-secure side panel and reload
Chemcassette
Note:
Remove and clean one port at a time to insure
proper orientation of tubing. Do not remove the capil-
lary tubing (microtubes).
CAUTION
Failure to replace and retighten hardware after service can
adversely affect monitor performance and electromagnetic
radiation compliance (EMC). Make certain all fasteners are
reinstalled and firmly fastened.
5-15
®
®
Remove these tubes one
at a time for cleaning