Honeywell Vertex M Technical Handbook page 152

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List
Number
Desc. (Short)
209
GATE TIMEOUT Gate Motor
210
OPTICS NOISE Optics Noise
211
OPTICS
FAILURE
212
GAS TABLE
ERROR
215
HIGH PYRO
FLOW
216
OPTICS DRIVE
LOW
217
OPTICS DRIVE
HI
Section 6 - Troubleshooting
Event History
Data Field
Description
0.0
Timeout
Counts
Optics Failure
Counts
Gas Table Error
High Pyrolyzer
flow cc/min.
Flow
Optics Drive Low Drive counts
Optics Drive High Drive counts
Problem
Gate motor timeout 7
Gate position sensor not
seconds
activated before timeout
Motor does not operate
Bad sensor or cable
Optics Noise
Poor grounding
Optics block cover loose
Count values from
Cable disconnected
optics reading
exceeded limits
Optics board defective
Sensor interface defective Contact Honeywell
Optics LED not properly
calibrated
Gas table corrupt or
No configuration loaded
missing
Proportional valve
Debris in valve
failure
Improperly
Change in flow system
characterized
configuration
proportional valve
Static flow reading
Transducer failure
Drive value below
Problem in optics block
allowable threshold
or sensor interface
electronics
Drive value above
Problem in optics block
allowable threshold
or sensor interface
electronics
6-12
Possible Cause
Check motor operation
using Maintenance/
Analyzer Operations/
Open Gate or Close Gate
Check motor connections
to sensor interface PCB
in analyzer
Check sensor connection
on PCB
Contact Honeywell
service
Contact Honeywell
service
Retighten or reinstall as
required
Check cable
Contact Honeywell
service
Contact Honeywell
service
service
Perform Load CC
Operation to recalibrate
Reinstall Configuration
Profile
Replace valve
Perform Auto Balance
operation
Service or replace
analyzer
Service or replace
analyzer
Service or replace
analyzer
Recovery