Honeywell Vertex M Technical Handbook page 114

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Vertex M
24-Point Continuous Monitor
TM
Auto Balance the Flow Rate
Note:
All analyzers should be in Monitor except for the one
being Auto Balanced.
Perform an Auto Balance for the following reasons:
When a line length is changed
If end-of-line filters are replaced
The gas family changes
When a new Chemcassette
When analyzers are added to a Vertex M rack
To perform an Auto Balance:
1. Make sure Chemcassette
2. Set analyzer to idle (not monitoring). See
Section 4.6.1 Run Time Options
procedure.
3. If not already open, touch Main Screen, Menu,
Calibration. The Flow Diagnostic window
opens.
4. Choose analyzer from the selection pad in the
upper right-hand corner.
5. Touch the Auto Balance function button. Vertex
M will characterize the flow between minimum
and maximum. This data will be used to set
the flow to 180 cc/min. when you return the
analyzer to monitor mode. Autobalancing takes
approximately 130 seconds to complete.
6. If required, repeat for other analyzers.
Note:
If Vertex M cannot adjust the flow to the proper range
(150 to 210 cc per minute at worst case condition)
consult Honeywell Analytics service.
Section 4 - Operation
is installed
®
is installed.
®
for the
4-38