Event
Event List
Number
Desc. (Short)
233
FAIL OPT
VERIFY
234
GRAY FAILED
235
AZ SW VER
WRONG
237
AZ VACUUM
LOW
238
OP SW
MISMATCH
239
OPTICS COMM
FAIL
240
OPTICS REF
FAIL
Section 6 - Troubleshooting
Event History
Data Field
Description
Optics
Adj Cnts
Verification
Failed
Light Gray Read
Optics Cnts
Failed
Analyzer
0
software version
mismatch
Inadequate
Observed
Analyzer Vacuum
vacuum in in
Hg
Optics block
0
software
mismatch.
Optics block
Error code
communication
failure
Optics reference
Reference
limit exceeded
value signal
Problem
Optics Verification
Chemcassette leader
Failed
not tight or improperly
positioned during
light gray to dark gray
calibration
Bad RFID tag
Dirty optics block
Bad optics PCB set
Optics values of this
Chemcassette leader
field were outside limits
not tight or improperly
positioned during
light gray to dark gray
calibration
Bad RFID tag
Dirty optics block
Bad optics PCB set
Software mismatch
Individual analyzer
Fault 219 has triggered.
observes inadequate
vacuum, stops
Circular tubing harness
monitoring
disconnected
Tubing pinched
The two optics blocks
Program update done
are not running the same
incorrectly.
software version.
Analyzer is not able to
Cable disconnected.
communicate with the
Software anomaly.
optics block.
Defective optics block.
The signal detected at
Degraded optics block.
the reference photodiode
is out of range.
6-15
Possible Cause
Reload Chemcassette
Load new Chemcassette
Clean and recalibrate
Service or replace
analyzer
Reload Chemcassette
Load new Chemcassette
Clean and recalibrate
Service or replace
analyzer
Reload program to all
analyzers
Contact Honeywell
service
Recover from Fault 219,
then reset Fault 237
Connect circular tubing
harness
Correct tubing dress in
cable carrier
Contact Honeywell
Service
Reload program to both
optics blocks.
Check the 14 pin
cable connection on
the sensor board.
Cycle power to analyzer.
Contact Honeywell service.
Contact Honeywell service.
Recovery