Vertex M
24-Point Continuous Monitor
TM
5.7.2 Replace Supply
1. Insert new supply in housing.
2. Verify handle is on the left of the power supply.
3. Seat firmly into place.
4. Verify latch or locking knob has locked supply
in the housing.
5. Tug on handle to verify supply is firmly seated.
5.8 Clean the Touch Screen
Clean the touch screen display with a lightly moistened
towel. Do not spray cleaner directly onto the glass.
Excess liquid will run down the screen and interfere
with operation.
Reference your touch monitor manual for any additional
information.
Section 5 - Maintenance
5.9 PLC Module Battery Backup Check
1. For status of battery locate indicator as shown
in photo on the left.
2. If battery indicator is illuminated the battery
has failed. Contact HA Service for corrective
action.
Note:
Some serial communication cards (located in slot
2) do not have battery backup.
5-13