Honeywell Vertex M Technical Handbook page 146

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List Desc.
Number
(Short)
103
TIMEOUT IDLE
104
TIMEOUT PNT
DIS
105
TIMEOUT ALM
DIS
106
POS PNT
PRESSURE
108
FLOW ADJ ERR
HI
109
CC EXPIRED
Section 6 - Troubleshooting
Event History
Data Field
Description
Timeout Of
0.0
Monitor
Timeout Run-
0.0
time Point
Disable
Timeout Run-
0.0
time Alarm
Disable
Positive Point
Pressure in.
Pressure
Hg
High Flow After
cc/min.
Adjustment
Chemcassette
0.0
Passed
Expiration Date
Problem
Analyzer in IDLE, LOAD
Out of Monitor time limit
CC, FLOW
has been exceeded
User error
Time limit too short
Operator error
Point disabled longer
than time limit setting in
the Configuration Profile
Time limit too short
Operator error
Point disabled longer
than time limit setting in
the Configuration Profile
Time limit too short
Positive pressure
Point pressure above
between sampling
atmospheric pressure
location and monitor
while in idle
Transducer error
Calibration error
Defective sensor
Offset incorrect
Flow cannot be adjusted
Debris in proportional
below 210 cc/min.
valve
Debris or defective valve
Proportional valve held
open
Operator error
Chemcassette installed
past its expiration date
Chemcassette expired
Expiration date reached
6-6
Possible Cause
Press RESET to
reinitialize timer.
Enter Monitor using
Runtime Options
Return analyzer to
Monitor Mode
Change time limit in
Configuration Profile
Press RESET to
reinitialize timer
Restore point to Monitor
Mode
Change time limit in
Configuration Profile
Press RESET to
reinitialize timer
Restore point to Monitor
Mode
Change time limit in
Configuration Profile
Relieve/reduce pressure
Contact Honeywell
Analytics Service
Contact Honeywell
Analytics Service
Contact Honeywell
Analytics Service
Verify range of valve
Perform Auto Balance to
clear debris
Replace with
0235-1279
Replace Chemcassette
Replace Chemcassette