Remove Filters; Replace Filters - Honeywell Vertex M Technical Handbook

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Vertex M
24-Point Continuous Monitor
TM

5.3.2 Remove Filters

Figure 5-1. Analyzer Filter Door
Note location of Acid Scrubber Filter
Arrow must point downward
Figure 5-2. Vertex M Filter Magazine
1. Open filter compartment by rotating the latch
downwards and pulling handle down until door
is at a 90˚ angle to the analyzer.
2. Remove filter magazine by grasping the
magazine with thumb and middle finger and
pulling straight out.
Load Filters in Magazine
1. Remove and discard used filter elements.
Section 5 - Maintenance
2. Position new filters in magazine with the filter
flow direction arrow pointing down. Press into
place.

5.3.3 Replace Filters

Figure 5-3. Analyzer Filter Door
Arrows must point downward
Figure 5-4. Vertex M Filters
1. Insert magazine into filter compartment.
2. Note orientation of arrow on side of magazine.
3. Close compartment door.
5-4