5.4.2 Remove Analyzer

1. Set analyzer to idle in the Runtime Options window.
Touch Main Screen, Menu, Maintenance and then Analyzer Operation.
Select the analyzer which you will replace. Touch Release Analyzer.
2. Open the front door and turn off power for the analyzer (and pyrolyzer switch if you are replacing a
Pyrolyzer).
3. Pull analyzer out of cabinet. Locks on the slides limit the analyzer travel.
4. Reach behind analyzer and remove four electrical cables.
5. Unlock and remove tubing harness.
6. Unlatch the slides.
7. Support the analyzer and remove it from the slides.

5.4.3 Install Analyzers

1. Reach into the slot and position cables out of the way.
2. Place analyzer on slides. Push in until locks engage.
3. Slide Analyzer out until fully extended.
4. Reach behind the analyzer; connect and lock the tubing harness.
Note:
When reconnecting the analyzer, connect the Analyzer Communications cable to the analyzer first.
5. Connect the Analyzer Communications, 24V Power Supply, Multifunction Connector and ChemCam
cables.
6. Carefully push analyzer partially into the cabinet and then pull out to verify all cables move freely and
the slides lock. Repeat in and out action to loosen slide. Push analyzer into cabinet.
Return to Service
1. Open the front door and turn analyzer power switch (and pyrolyzer power switch if neccessary) on.
2. Re-install Configuration Profile.
3. Install Chemcassette
4. Return analyzer to monitor mode in Runtime Options Menu.
Section 5 - Maintenance
.
®
5-7