Vertex M
24-Point Continuous Monitor
TM
Choose the analyzer from the selection pad in the upper right-hand corner. Analyzer Information lists the following:
•
Profile ID - The Profile ID is a unique number generated whenever the Vertex M Profile Management
utility saves a configuration file. Vertex M stores the ID number in the DAq, the PLC and each analyzer.
If the profile ID numbers do not match, a fault is generated and monitoring will not start.
•
Gas family information
•
Network IP address
•
Analyzer software version number
•
Analyzer Serial Number
•
Time remaining on filter
•
Time remaining on Chemcassette
•
Data logging settings for the analyzer
•
Pyrolyzer analyzers will also display the pyrolyzer temperature.
Section 4 - Operation
tape
®
4-42