5.3.4 Change Chemcassette
Change the Vertex M Chemcassette
the following reasons:
Scheduled end-of-tape service
Low Chemcassette
Chemcassette
®
End of Chemcassette
Transport error
Preparation
1. Reset End of Chemcassette
2. Touch Main Screen, Menu and then
Maintenance.
3. Touch the module number in the Maintenance
window.
4. Touch Load CC. Follow the on-screen
instructions which will guide you through the
following sequence.
Change Chemcassette
1. Pull analyzer out of cabinet.
2. Remove the old Chemcassette
3. Install new tape. Verify the RFID tag is aligned
with the RF sensor.
Touch the NEXT button to read RF sensor.
4. Thread tape leader.
Section 5 - Maintenance
Tape
®
tape for any of
®
warning (fault 102)
®
has expired (fault 109)
(fault 203)
®
event if present.
®
Tape
®
.
®
5. Position leading alignment mark on
Chemcassette
head.
6. Touch the NEXT button to verify optics. The
tape advances as the Vertex M verifies the
optics.
7. Push the analyzer into the cabinet.