Inspection and
Maintenance

Inspection and Maintenance

WARNING !
To avoid any risk of infection, clean and disin-
fect incubator and accessories before any
maintenance according to established hospital
procedures – this applies also when returning
units or parts for repair.
WARNING !
Always disconnect supply before servicing.
WARNING !
Never operate the incubator if it has suffered
physical damage or does not seem to operate
properly. In this case always refer servicing to
properly trained and factory authorized service
personnel.

Replacement of Parts

Air filter
replace after two months, see
page 30.
Discard with regular waste.
Hand port seals
replace, when the material
becomes brittle or sticky or when a
proper seal with the port can no
longer be obtained.
U-grommets
replace, when the material becomes
brittle or sticky or when segments
are missing.
Fan motor
lubricate with 10 drops of
51524-HLP 32 oil every six months
by trained service personnel.
NiCd battery for
to be replaced once a year by
power failure alarm factory trained technician
WARNING !
Treatment of batteries
– Do not throw into fire! risk of explosion
– Do not force open! cells contain
corrosive acid
60
Inspection and
Maintenance
Replacement of Parts