Hose Connections; Temperature Control Unit (With Liquid) - Thermo Scientific HAAKE RotoVisco 1 Instruction Manual

Rheometer
Table of Contents

Advertisement

Installation

8.3 Hose connections

8.3.1 Temperature control unit (with liquid):

The temperature control unit with liquid temperature
control is connected with hoses to a heating bath and
circulator.
!
While fastening the hose connections at the con-
necting nozzles (14) of the temperature control unit
the connecting nozzles must be held up by
wrenches.
!
Direction of flow of the liquid
Temperature range up to 150 C:
Viton hoses
Hoes nozzle -Set
Temperature range of 100 C up to 350 C:
Metal hoses
Hoes nozzle -Set
Liquid temperature control:
Temperature range up to 100 C:
Thermal liquid in the temperature range between --50
C and 30 C: water with anti-freeze.
Temperature range from 5 C to 90 C: distilled water.
Distilled water
Normal mains water causes cable deposits and
frequently requires the unit to be descaled.
In principle, water up to 95 C can be used. However,
above 80 C, so much water evaporates that it requires
frequent topping--up.
Water with anti- -freeze
If you intend working at temperatures below 5 C,
anti--freeze must be added to the water. The added
amount of anti--freeze should be sufficient for a
temperature that is about 10C lower than the intended
working temperature. This prevents the freezing of
water on the evaporator coil of the cooling circuit, the
surface of which is always much colder than the
23
222--1492
333-0294
222--1492