10. Operating

10.1 Switching on

When all connections have been established and the
1
supply lines are active, switch--on the mains switch 6:
 The green display 4 on the measuring table of the
rheometer indicates the ready state for operation.
The following operation sequence should be used to be able
to make the measurements at short notice:
-- Instrument:
Automatic device initialization
Raises the lift. Never switch off the unit during lift move-
ment.
!
In case of loss of power during the lift movement,
the unit has to be initialized again.
010
Then the following display appears:
-- Operator: Insert sensor if necessary
-- Instrument: Lowers the lift for zero point determination.
Status
!
Due to the rotation of the pin spots can appear on
the measuring plate. This is normal and does not
constitute any damage to the unit.
Then move the lift again to cleaning position (i.e. up), the in-
strument immediately shows the MENU selection window
(A) for defining, if necessary, the operator, sample name etc.
010

10.2 Working with the display unit

In MENU activate CONFIGURATION by pressing the key.
Status
Pressing the key leads directly into the submenu.
100
Lift
Menu
Status
39