Flat Sieve - Thermo Scientific HAAKE RotoVisco 1 Instruction Manual

Rheometer
Table of Contents

Advertisement

Pflege / Wartung

15.4 Flat sieve

To keep dirt out of the tempering unit there is a flat sieve
(003--5266) inserted in the suction side 14 (IN). This sieve
has to be changed or at least cleaned at the following inter-
vals:
every six months, or
as soon as the through--flow is too small.
Changing the flat sieve:
Switch the unit off.
1
Separate the hose from the hose connectors (IN)
2
!
When loosening the hose connections at the tem-
pering unit, hold the connectors with an open- -jaw
spanner
Loosen and remove the hose connections from the
3
tempering unit
Unscrew the flat sieve from the connector with a
4
wrench.
Clean the flat sieve or replace it with a new one.
5
Insert the flat sieve and fasten it to the hose connec-
6
tions together again
Reconnect the hoses
7
!
Tighten all the hose connections with hose
clamps.
142
14