Sensor Systems
Measuring Ranges
The different specifications of the models HAAKE RotoVisco
1 and HAAKE RheoStress 1 effect the measuring ranges.
For the values of measuring ranges please see chapter
"Technical Specifications".
The measuring range can be illustrated in a diagram where
.
= f ( γ
) and = f ( γ
- Range
The shear stress measuring range results from the mea-
surement geometry and the presetting range of the torque
Md. Now it is quite simple to estimate the smallest and larg-
est shear stress value for a sensor system by using the fol-
lowing calculations:
= Md
min
(min)
= Md
max
(max)
The limit values of the diagrams differ according to the model
specific torque values from the above table.
.
γ
- Range
Similar to the shear stress range there is also a meaningful
measuring range for the shear rate γ
tions can be derived:
.
γ
=
min
(min)
.
γ
=
max
(max)
The limit values of the diagrams differ according to the model
specific torque values from the above table.