Plumbing Connections - Thermo Scientific 2210AM Manual

Ammonia analyzer
Table of Contents

Advertisement

Figure 4. Typical external plumbing connections
1. It is recommended that you install an
external sample shut off valve to allow
control while accessing or service to the
sample chamber.
2. Air purging of upper and lower
NOTES:
enclosures is recommended.
3. Each analyzer MUST have its own drain
tube vented to atmosphere.
4. The installation of a power circuit
breaker is highly recommended.

Plumbing Connections

Sample inlet, sample return and drain connections are made
on the bottom side of the instrument. The sample inlet and
sample return fittings on the bottom of analyzer require ¼" OD
(6 mm OD) flexible tubing. The drain fitting at the bottom of
the analyzer requires
" OD (9.6 mm OD) flexible tubing. The
air purge inlets require ¼" OD tubing. The air purge supply
must be filtered externally and free of oil and any other
particulates.
The ⅜" (9.6 mm) DRAIN MUST BE
VENTED TO ATMOSPHERE. The
sample drain contains analysis wastes,
which include both sample and
chemical reagents. Please check with
DANGER
local authorities concerning proper
disposal of waste from this product.
10 |
Installation Environmental Considerations
The enclosure drain produces no waste
under normal use. However, in the event
of spills or leaks, some waste may be
produced.
NOTES:
The analyzer is shipped with all of the
internal tubing attached. Before
starting up the instrument, verify that the
tubing lines are secured.
Figure 5. Internal tubing and tube routing
interconnections
Installing Peristaltic Pump Head
Tubing
Press the STOP key.
Disengage the U shape pressure plate of the pump head
assembly by pressing with two fingers and gently move
upside as shown in Figure 6.
Replace the old tube by sliding off the barbed fitting. If
necessary, clean the pump rollers.
Lubricate the tubing with a pea-sized quantity of silicone oil.
Fit the new tube and replace in the pressure plate. Note that
the pressure plate sides must click properly in the pump head
assembly and align on both sides.
Repeat above steps for all pumps.
You can test the pump under Diagnostics menu for proper
operation:
MENU
SERVICES
STIRRER
DIAGNOSTICS
PUMPS &