Unit Description

7.3 Temperature control:

!
For applications at elevated temperatures above 250C
it is a must to switch the fan for the air bearing cooling
on stage 2 (on the rear) and use ceramic shafts only.
Furthermore use the cone heater TC1 as thermal shield
to prevent the heating up of the air bearing housing!
TCO
Glass plate with a Pt100 temperature sensor
(not temperature controlled).
TCL/Z
Liquid temperature controlled receptacle for cylinder
measuring systems with direct contact of the thermal
fluid from --40 up to 200C.
TCL/P
Liquid temperature controlled measuring plate for
parallel plate or cone & plate measurements (with
external circulator from --20 up to 300C).
TCP/P
Measuring plate with Peltier temperature control for
TCP/PE
parallel plate or cone & plate measurements from
--40 up to 180C (external heat transfer required).
TCE/P
Electrically temperature controlled measuring plate for
parallel plate or cone & plate measurements from
--20C up to 350C .
TCE/PC
Electrically temperature controlled measuring plate for
parallel plate or cone & plate measurements with
electrical cone heating from --40C up to 350C.
SHRP
Special temperature controlled system utilizing
thermal liquid as standardized for asphalts and
bitumen.
Temperature range: 0C (--25C)... +90C
(for HAAKE RheoStress 1)