12.3 Cylinder Sensor System Z

Application
This sensor system is preferred for medium viscous liquids
when a comparability of the measurements in accordance
with DIN 53018 is requested. These sensors have an ex-
tremely small front surface influence and are therefore in-
tended for exact measurements. Temperature programs are
not recommended as the volume of the trapped air bubble
will change its volume with the temperature. This also has an
effect on the front surface influence.
Sensor System Z
The sensor system Z comprises a collapsible beaker Z43
and 3 rotors with different radii.
222-1488
222-1461
222-1460
222-1459
Z43
Z31
Z38
Z41
Sealing set for Z40
Order no. 222--1290