Double Gap Cylinder Sensor Dg43 According To Din 53544 - Thermo Scientific HAAKE RotoVisco 1 Instruction Manual

Rheometer
Table of Contents

Advertisement

Sensor Systems
12.4 Double Gap Cylinder Sensor DG43
according to DIN 53544
Application
This sensor system is preferred for low viscous liquids
(<1000 mPa  s) or for small sample volumes. The double
shearing surfaces of this particular system result in a higher
shear stress than what is customary for comparable DIN
sensors. It is standardized as DIN 53544 for measurements
with low viscous glues.
Sensor System DG43
The sensor system DG43 is made up of a dismountable bea-
ker and either the bell-shaped rotor.
The temperature vessel should not be employed for temper-
atures above 200C. There must be a sufficiently long heat-
ing period so that the inner part of the beaker, which is not
temperature controlled, can adopt the desired temperature.
Heating times of 5 to 20 minutes, depending on the viscosity
and liquid, are quite common.
93
Rotor DG43
222-1559
002-6748
Beaker DG43
222-1489
003-4727
Sealling set for DG41/43
222--1293