Plumbing Connections - Thermo Scientific Orion 2230XP User Manual

Silica analyzer
Table of Contents

Advertisement

Figure 3: Typical external plumbing
connections
2230XP Silica Analyzer
INSTALL HEIGHT TO
OPERATOR EYE LEVEL
FLOW DIRECTION
FILTERED AND OIL FREE
AIR PURGE SUPPLY LINE
FLOW DIRECTION
CONNECT TO
ATMOSPHERIC DRAIN
1. It is recommended that you install an external sample shut off valve to
allow control while accessing or service to the sample chamber.
2. Air purging of upper and lower enclosures is recommended.
NOTES:
3. Each analyzer MUST have its own drain tube vented to atmosphere.
4. The installation of a power circuit breaker is highly recommended.

Plumbing Connections

Sample inlet, sample return and drain connections are made on the bottom side of the
instrument. The sample inlet and sample return fittings on the bottom of analyzer require ¼-in
OD (6 mm OD) flexible tubing. The drain fitting at the bottom of the analyzer requires ⅜ -in OD
(9.6 mm OD) flexible tubing. The air purge inlets require ¼-in OD tubing. The air purge supply
must be filtered externally and free of oil and any other particulates.
Installation Environmental Considerations
RGT1
RGT3
SAMPLE
(BACK) N.C.
STANDARD (FRONT) N.O.
RGT2
RGT3
RGT1
STD
SAMPLE CHAMBER
DRAIN
SAMPLE (INLET)
USER TO INSTALL A 40 to 60 MICRON
FILTER, SIZE IS BASED ON SAMPLE
CONDITION, SET FLOW RATE
BETWEEN : 50 TO 500 ML/MIN
Section 3
AC MAINS
1
AC POWER
0
POWER CIRCUIT BREAKER
MUST BE CLEARLY IDENTIFIED
SHUT-OFF
VALVE
SAMPLE IN
PROCESS LINE
SAMPLE RETURN
Thermo Scientific Orion
19