Temperature Control Units

11.3 Temperature control unit TCL/P

The temperature control unit has a holding for the measuring
plate. The measuring plate is put on the thermal liquid heated
temperature control unit and is fastened.
Cones and measuring plates should have the same diame-
ter e.g. as cone C60/1 and the measuring plate MP60 or
plate PP35 and measuring plate MP35. If not, huge measur-
ing faults are the consequence !
For measurements at temperatures higher than 70C it is
recommendable to use at least the sample protection shield
(222-0608) to reduce the loss of heat. In any case, the ap-
plication of measuring cones and measuring plates with a
ceramic shaft that do carry off less heat is recommendable.
The sensor system and the temperature control unit need
cooling that is dependent on the load and the temperatures
during the measurement.
!
While fastening the hose connections at the con-
necting nozzles(14) of the temperature control
unit::
- - ensure the correct flow direction (IN/OUT)
- - the connecting nozzles must be held secured
using a wrench.
For the cooling of the temperature control unit hoses (with
 8mm) can be connected to a liquid circulator.
maximum conduction pressure 0.5bar (water circulation).
59