Maintenance Instructions

15.2 Maintenance Instructions HAAKE Rotovisco1

The system check on the HAAKE Rotovisco1 must be
carried out once yearly.
1.
Check the lifting device with drive and guideway
cover. *
2.
Clean and lubricate screw.*
3.
Check safety device and final cut--off lift.*
4.
Test and adjust surface plate and plate carrier
mechanically.*
5.
Update devices to newest operating software.
6.
Check speed step with digital speed measurem
ent device.*
7.
Calibrate torque measurement with thread measu
ring instrument and box of weights, check linearity of
torque indication.*
8.
Test measurement with normal liquid in CR mode.
9.
General function test.*
10. Check all sensors for smooth running and damage.
11. Check temperature stabiliser (if applicable):*
Liquid tempering:
Check hose connections, sensors, update operatin
software.
TCE/P electric tempering:
Check heating elements, sensors, tempering.
TCP/P and TCP/PE Peltier tempering:
Check tempering, safety circuit, tempering.
* Check by official Thermo Fisher Scientific service.
Points 8 and 10 are recommended every 3 months.