Temperature Control Units

11.4 Temperature control unit TCE/P

The temperature control unit has holding a for the measuring
plate. The measuring plate is put on the thermal liquid heated
temperature control unit and is fastened.
Cones and measuring plates should have the same diame-
ter e.g. as cone C60/1 and the measuring plate MPC60 or
plate PP35 and measuring plate MPC35. If not, huge mea-
suring faults are the consequence !
For measurements at temperatures higher than 80C it is
recommendable to use at least the sample protection shield
(222-0608) to reduce the loss of heat. In any case, the ap-
plication of measuring cones and measuring plates with a
ceramic shaft that do carry off less heat is recommendable.
The temperature control unit is connected to a liquid circula-
tor via hoses ( 8mm) .
For the TCE/P measuring unit and the open--bath and heat-
ing circulators the necessary tubing hoses are not part of the
standard accessories. They have to be ordered separately.
!
The cooling (liquid or gas) is controlled directly in
the measuring device via the built-in solenoid
valve according to the requirements. It is important
that the cooling medium used is selected in the
RheoWin software, as the control parameters (PID)
are thus selected.
The switching condition of the valve is „open".
Utilizing the optional electrical cone heating extends the
temperature range up to 350C.