Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series)
(cont.)
Sample acquisition mode,
typical
Delta voltage measurement
between any two samples
acquired with the same setup
and ambient conditions,
typical
Position range
6
±(DC Gain Accuracy × | reading -(net offset) | + offset accuracy + 0.15 division + 0.6 mV)
±(DC Gain Accuracy × | reading | + 0.15 division + 1.2 mV)
Convert offset, position and the constant offset term to volts by multiplying by the appropriate volts/
div.
Specification applies to any sample and to the High, Low, Max, Min, Mean, Cycle Mean, RMS, and
Cycle RMS measurements. Delta volts specification applies to subtractive calculations involving
two of these measurements. Delta volts specification applies to the Positive Overshoot, Negative
Overshoot, peak-peak, and amplitude measurements.
± 5 divisions