Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
Ch
Scale
setting
Ch2
100 mV
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
Termination
Position
Offset
setting
setting
setting
(Divs)
1.5 V
0
2.5 V
0
1.5 V
0
0.5 V
3.4 V
0
3.4 V
0
2.4 V
-3.4 V
0
-3.4 V
0
-2.4 V
-1.5 V
0
-2.5 V
0
-1.5 V
0
-0.5 V
0 V
0
0 V
0
+1 V
0
-1 V
1.5 V
0
2.5 V
0
1.5 V
0
0.5 V
3.4 V
0
3.4 V
0
2.4 V
Generat
Measurement
Difference of
or
mean
measurement
setting
means
2.65 V
2.35 V
1.65 V
1.35 V
650 mV
350 mV
3.55 V
3.25 V
2.55 V
2.25 V
-3.25 V
-3.55 V
-2.25 V
-2.55 V
-2.35 V
-2.65 V
-1.35 V
-1.65 V
-350 mV
-650 mV
300 mV
-300 mV
1.3 V
700 mV
-700 mV
-1.3 V
2.8 V
2.2 V
1.8 V
1.2 V
800 mV
200 mV
3.70 V
3.10 V
2.70 V
2.10 V
Accuracy limits
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+294 mV to
+306 mV
+588 mV to
+612 mV
+588 mV to
+612 mV
+588 mV to
+612 mV
+588 mV to
+612 mV
+588 mV to
+612 mV
+588 mV to
+612 mV
+588 mV to
+612 mV
+588 mV to
+612 mV