Warning: The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified
to do so. Refer to all safety summaries prior to performing service.
Mso2000b series; mso2000 series; dpo2000b series; dpo2000 series (441 pages)
Summary of Contents for Tektronix 6 Series
Page 1
6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions (MSO64B, MSO66B, MSO68B) Warning: The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Sanitizing the SSD................................. 9 Troubleshooting ..................................10 How to sanitize a non-functional instrument........................10 Repair charges..................................10 Change log ....................................11 6 Series B MSO Declassification and Security Instructions document change log..............11 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
List of Tables List of Tables Table 1: Volatile memory................................1 Table 2: Non-user-accessible memory............................2 Table 3: Media and Data export devices ............................7 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Instrument Declassification. A term that refers to procedures that must be undertaken before an instrument can be removed from a secure environment. Declassification procedures include memory sanitization and memory removal, and sometimes both. 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Page 6
Preface 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
≥512 MB graphics operations from the data instrument instrument for a for a minimum of minimum 30 seconds. of 30 seconds. Table continued… 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Page 8
9 trap door in on the Holds user- the bottom bottom of storable of the data such instrument. instrument. waveforms, measureme nt results, instrument settings. Table continued… 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Page 9
≥64 Kbit does not does not configuratio contain user contain user data or data or settings. settings. Clearing Sanitizing would would disable disable instrument instrument functionality. functionality. Table continued… 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Page 10
Clearing Sanitizing embedded would would controller disable disable firmware. instrument instrument functionality. functionality. Ethernet address is stored in this device. Table continued… 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Page 11
Clearing Sanitizing would would disable disable instrument instrument functionality. functionality. Table continued… 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Page 12
≥0.33 Mbit FPGA FPGA on contain user contain user configuratio data or data or processor settings. settings. carrier Clearing Sanitizing board would would disable disable instrument instrument functionality. functionality. 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
Option 6-SEC is installed. Option 6-SEC for a secure instrument Option 6-SEC provides the highest level of instrument security for 6 Series B MSO products. Option 6-SEC features include: • Password protection to enable/disable external communication ports •...
5. Recall the file from the Recall Setup dialog box (File > Recall > Setup tab). For more information on using programmatic commands, refer to your product programmer manual, available at www.tek.com/manuals. 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
SSD retain their data and files after running SQE Tests TekSecure. Sanitizing the SSD The instrument does not have any function to sanitize the entire SSD and retain instrument operation. 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...
How to sanitize a non-functional instrument If your instrument is not functioning, proceed as follows to sanitize the instrument to return to Tektronix for repair: 1. Remove any attached USB flash drives or external USB drives from your oscilloscope. Refer to your company's internal policies regarding handling or disposal of the flash drives.
Change log Change log 6 Series B MSO Declassification and Security Instructions document change log Document part Revision Change description number date 6 Series B Mixed Signal Oscilloscopes Declassification and Security Instructions...