Tektronix MSO70000C/DX Technical Reference page 226

Series oscilloscopes specifications and performance verification
Hide thumbs Also See for MSO70000C/DX:
Table of Contents

Advertisement

Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
Channel Scale
setting
100 mV
200 mV
500 mV
Ch2
10 mV
Ch2
20 mV
50 mV
214
Position
Offset
Generator
setting
setting
setting
(Divs)
0
0 V
+380 mV
-380 mV
-5
+2.0 V
+2.88 V
+2.12 V
5
-2.0 V
-2.12 V
-2.88 V
0
0 V
+760 mV
-760 mV
-5
+1.25 V
+3.01 V
+1.49 V
5
-1.25 V
-1.49 V
-3.01 V
0
0 V
+1.90 V
-1.90 V
-5
0 V
+4.40 V
+0.60 V
5
0 V
-0.60 V
–4.40 V
0
0 V
+38.0 mV
-38.0 mV
-5
+0.45 V
+538 mV
+462 mV
5
-0.45 V
-462 mV
-538 mV
0
0 V
+76.0 mV
-76.0 mV
-5
+0.4 V
+576.0 mV
+424.0 mV
5
-0.4 V
-424.0 mV
-576.0 mV
0
0 V
+190 mV
-190 mV
-5
+0.25 V
+690 mV
+310 mV
5
-0.25 V
-310 mV
-690 mV
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
Measurement
Difference of
mean
measurement
means
Accuracy limits
+744.80 mV to +775.20 mV
+744.80 mV to +775.20 mV
+744.80 mV to +775.20 mV
+1.49 V to +1.55 V
+1.49 V to +1.55 V
+1.49 V to +1.55 V
+3.724 V to +3.876 V
+3.724 V to +3.876 V
+3.724 V to +3.876 V
+74.48 mV to +77.52 mV
+74.48 mV to +77.52 mV
+74.48 mV to +77.52 mV
+148.960 mV to +155.040 mV
+148.960 mV to +155.040 mV
+148.960 mV to +155.040 mV
+372.40 mV to +387.60 mV
+372.40 mV to +387.60 mV
+372.40 mV to +387.60 mV

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MSO70000C/DX and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents