DC Offset Accuracy
Performance checks Vertical scale
Channel 3 DC Offset
1 mV/div
Accuracy, 20 MHz
1 mV/div
BW, 1 MΩ
100 mV/div
100 mV/div
500 mV/div
500 mV/div
1.01 V/div
1.01 V/div
3 V/div
3 V/div
Channel 4 DC Offset
1 mV/div
Accuracy, 20 MHz
1 mV/div
BW, 50 Ω
100 mV/div
100 mV/div
Channel 4 DC Offset
1 mV/div
Accuracy, 20 MHz
1 mV/div
BW, 1 MΩ
100 mV/div
100 mV/div
500 mV/div
500 mV/div
1.01 V/div
1.01 V/div
3 V/div
3 V/div
1
Use this value for both the calibrator output and the oscilloscope offset setting.
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series
Performance verification (MSO/DPO5000/B series)
1
Vertical offset
Low limit
900 mV
895.3 mV
-900 mV
-904.7 mV
9.0 V
8.935 V
-9.0 V
-9.065 V
9.0 V
8.855 V
- 9.0 V
-9.145 V
99.5 V
98.80 V
-99.5 V
-100.2 V
99.5 V
98.4 V
-99.5 V
-100.6 V
900 mV
895.3 mV
-900 mV
-904.7 mV
5.0 V
4.965 V
-5.0 V
-5.035 V
900 mV
895.3 mV
-900 mV
-904.7 mV
9.0 V
8.935 V
-9.0 V
-9.065 V
9.0 V
8.855 V
- 9.0 V
-9.145 V
99.5 V
98.80 V
-99.5 V
-100.2 V
99.5 V
98.4 V
-99.5 V
-100.6 V
Test result
High limit
904.7 mV
-895.3 mV
9.065 V
-8.935 V
9.145 V
-8.855 V
100.2 V
-98.80 V
100.6 V
-98.4 V
904.7 mV
-895.3 mV
5.035 V
-4.965 V
904.7 mV
-895.3 mV
9.065 V
-8.935 V
9.145 V
-8.855 V
100.2 V
-98.80 V
100.6 V
-98.4 V