Example Acceleration and Deceleration Settings

The values associated with each acceleration and deceleration setting are given in the table below.
Table 6.1 Acceleration and Deceleration Profiles
Setting
Maximum acceleration (not available
with the JS-24.38 and JS-24.15 rotors)
Timed acceleration
Maximum deceleration
Timed deceleration
Coast

Login Page

Figure 6.10 Login Page
If the user login requirement has been enabled for the instrument, you must log in to the system
use it. Select the
PN B38322AA
Description
Full torque is used from 0 RPM until set speed is reached.
Acceleration time from 0 to 500 RPM can be set from 1 to 10
minutes. Above 500 RPM, full torque is used until set speed is
reached.
Full brake from set speed to near 0 RPM. Reduce brake is used
during the last few RPM until 0 is reached to minimize sample
disturbance.
Full brake from set speed to 500 RPM. From 500 to 0 RPM,
deceleration time can be set from 1 to 10 minutes.
No brake is used. Rotor coasts from set speed to 0 RPM. The time
required for a rotor to come to a complete stop can vary from 1 to
9 hours depending upon the set speed while in operation and the
rotor's mass.
button on the
Home Page
Login
to display the
page.
Login
6
Functional Pages

Login Page

6-13