Entering Acceleration And Deceleration Rates - Beckman Coulter Avanti J-E Instruction Manual

Table of Contents

Advertisement

Operation

ENTERING ACCELERATION AND DECELERATION RATES

Action
1. Press [
ACCEL
2. Press the keypad number for
the required acceleration
rate—1 = MAX, 2 = SLOW.
3. Press the next run-parameter
key or [
ENTER
4. Press [
DECEL
5. Press the keypad number for
the required deceleration
rate—1 = MAX; 2 = SLOW;
3 = OFF.
3-8
The instrument provides a choice of two acceleration rates and three
deceleration rates to protect the gradient and sample-to-gradient inter-
face. The acceleration time is the time it takes a rotor to reach set
speed from rest. The deceleration time is the time it takes a rotor to
decelerate from set speed to rest.
Result
].
Acceleration options are displayed.
Entered acceleration rate appears on the display.
].
].
Deceleration options are displayed.
Entered deceleration rate appears on the display.