Preinstallation Requirements

Preinstallation Requirements
Table A.1 Nominal Supply Voltage Ratings for the Avanti JXN Series
Nominal
Instrument
Voltage Rating
Single-phase,
200/208/240 V
Single-phase,
230 V
Three-phase,
220/380 V +
a
Neutral
a. Unbalanced three-phase. Split for single-phase operation internally.
To ensure optimal safety, the centrifuge should be wired to a remote emergency switch (preferably
outside the room where the centrifuge is housed, or adjacent to the exit from that room). In case of
a malfunction, the centrifuge can be disconnected from the main power source by removing the
Mains (power) plug from the outlet receptacle.
WARNING
To reduce the risk of electrical shock, this equipment uses a three-wire or five-wire
electrical cord and plug to connect the centrifuge to earth-ground. To preserve
this safety feature, make sure that the matching wall outlet receptacle is properly
wired and earth-grounded.
Prior to instrument purchase, the power configuration should be determined and the appropriate
instrument ordered.

Single and Three-Phase Power Connections

Figure A.1
shows the power connection for single-phase centrifuges, including earth-ground and
two power leads with 30-ampere circuit breakers.
three-phase centrifuges, including earth-ground and three power leads with minimum 16-ampere
circuit breakers and common neutral connection.
A-2
Nominal Supply
Power Cord and Plug Description
Frequency
180–264 V,
permanently attached three-wire
50/60 Hz, 30 A
UL/CSA-approved cord with NEMA
type 6-30P plug
180–264 V,
permanently attached three-wire
50 Hz, 30 A
CENELEC harmonized cord without
plug
313–457 V,
permanently attached five-wire
plus neutral,
CENELEC harmonized cord without
50 Hz, 16 A
plug
Figure A.2
shows the power connection for
Table A.2
contains wiring information.
PN B38322AA