Table 1.1 Avanti JXN Series Specifications
Specifications
Speed
Time
Temperature
Accel/Decel
Dimensions
PN B38322AA
Description
Setting range
Speed control (1,000 to
10,000 RPM)
Speed control (10,001 to
max speed)
Setting range
Actual display
2
ω
t setting range
2
ω
t actual display
Setting range
Accuracy
Ambient temperature range
Over temperature shutdown
c
point
Cooling fluid
Acceleration
Deceleration
Width
Depth
Height (with door closed)
Height (to top of control
head)
Height (to top of open door)
Weight
Avanti JXN-26
500 to 26,000 RPM (in 10
RPM increments) or
equivalent RCF (in 10 x g
increments)
±10 RPM
±0.1%
1 min to 99 hr 59 min, ω
Indicates time remaining (timed run) , ω
time (hold run)
12
Up to 2.6682 x 10
radians squared per
second
Indicates accumulated centrifugal effect to five
significant digits (in exponential notation)
-10 to +40°C (in 1°C
increments)
±2°C of chamber temperature (after equilibration)
16 to 38°C
50°C
Refrigerant 404A (HVC)
Maximum, timed (1 to 10 min from 0 to 500 RPM)
Maximum, timed (1 to 10 min from 500 to 0 RPM), or
coast
71 cm (28 in.)
86 cm (34 in.)
86 cm (34 in.)
120 cm (47.3 in.)
149 cm (58.5 in.)
290 kg (640 lb)
Description
Specifications
Avanti JXN-30
500 to 30,000 RPM (in 10
RPM increments) or
equivalent to RCF (in 10 x g
increments)
±0.15%
2
t, or continuous (hold)
2
t, or elapsed
12
Up to 3.5524 x 10
radians squared per
second
-20 to +40°C (in 1°C
increments)
a
b
55°C
141 cm (55.5 in.)
310 kg (680 lb)