Entering Acceleration And Deceleration Rates - Beckman Coulter Optima MAX-XP Instructions For Use Manual

Ultracentrifuge
Table of Contents

Advertisement

Operation
Manual Operation
1
On the Main screen, press the
• The
• The acceptable values for run temperature are shown above the display field: from
0 to 40°C.
2
Enter the temperature on the keypad, and press the
• The Main screen returns to view, and the set run temperature appears above the
Temp
NOTE
If you enter a temperature value greater than 40, the
3
To change the set temperature at any time during a manual run, repeat Steps 1 and 2.
• The ultracentrifuge accepts the new temperature immediately.
NOTE
If the rotor temperature is not within 3°C above the set temperature for more than 40 minutes,
a diagnostic message appears. If a run is in progress, the rotor will decelerate to a stop. See
CHAPTER 4, Troubleshooting

Entering Acceleration and Deceleration Rates

The ultracentrifuge provides ten acceleration rates and eleven deceleration rates to protect the
gradient and sample-to-gradient interface.
numbers on the touchscreen interface. Acceleration time is the time it takes a rotor to reach
5000 RPM from rest. At 5000 RPM, maximum acceleration takes over until the rotor reaches set
speed. Deceleration time is the time it takes a rotor to decelerate from 5000 to rest. From set speed
to 5000, the rotor decelerates with full dynamic braking.
3-16
TEMP
screen appears.
Enter Run Temperature
display.
.
button.
button to accept.
OK
button will be grayed out.
OK
Table 3.1
lists these rates by their corresponding
393552AH