Manual Operation

Entering Run Temperature

Run temperature can be set from 2 to +40°C. If no value is entered, the centrifuge selects the last
entered temperature.
1
Press
TEMP
• The cursor and the temperature range (2 to 40°C) appear.
2
Use the keypad or arrow keys to enter required temperature.
• Entered temperature appears on the display.
• (If the entered temperature is outside the valid temperature range, an error message
will appear and the input will be cleared.
a. Enter a valid temperature.)
3
Press the next run-parameter key or

Entering Acceleration and Deceleration Rates

You can select one of ten acceleration rates and eleven deceleration rates to protect the gradient
and sample-to-gradient interface. Select rates depending on the type of run you are performing—
for pelleting runs, where sample mixing is not a concern, maximum acceleration and deceleration
can be used. If running delicate gradients, a lower setting may be needed. Typical acceleration and
deceleration rates are listed in
accelerates and decelerates at maximum.
Table 2.1 Acceleration and Deceleration Rates (in Minutes:Seconds) to and from Maximum Speed
Rate
SX4750
(4300 RPM)
Accel
Decel
10 (MAX)
1:24
1:21
9
2:03
2:07
8
2:14
2:18
7
2:27
2:33
2-6
.
Table
2.1. If no rate is selected, the centrifuge automatically
SX4750
SX4750A
(4000 RPM)
(4300 RPM)
Accel
Decel
Accel
1:14
1:14
1:23
1:53
1:59
2:00
2:04
2:08
2:14
2:16
2:21
2:27
.
ENTER/SAVE
SX4750a
4000 RPM
Decel
Accel
1:21
1:14
2:06
1:50
2:19
2:04
2:30
2:17
a
FX6100
(10,200 RPM)
Decel
Accel
Decel
1:16
1:24
1:23
1:58
2:23
2:26
2:09
3:24
3:25
2:23
4:28
4:23
PN B07493AD