Audible Sounds

The Avanti JXN series instrument makes an audible sound for the following events:
• Boot up
• Start of Run
• End of Run
• Diagnostics/Alert
• Door open during Zonal mode
• Key Click
Sound volume can be regulated through the
through the System Options. See
The Door Open during Zonal mode sound cannot be changed or muted. It will play every 5 seconds
when the door could be open.

Available Rotors

Refer to the applicable rotor manual for complete rotor descriptions.
Table 1.2 Available Rotors
Rotor Profile
PN B38322AA
Set Sound
Custom Sounds
a
Description
JXN-26
Max
b
RPM
JA-30.50 Ti
26,000
Fixed Angle, 34°
(8 place)
r
= 108 mm
max
JA-25.50 Ti
25,000
Fixed Angle, 34°
(8 place)
r
= 108 mm
max
JA-25.15 Ti
25,000
Fixed Angle, 25°
(24 place)
r
= 106 mm
max
(outer row)
r
= 86 mm
max
(inner row)
Page, and some sounds can be customized
Page.
JXN-30
JXN-26
JXN-30
Max
Max
Max
b
RCF
RCF
RPM
(´ g)
(´ g)
30,000
81,770
108,860
25,000
75,600
75,600
25,000
74,200
74,200
60,200
60,200
1
Description

Available Rotors

Max
Rotor
Capacity
Manual
Number
8 ´ 50 mL
J-TB-070
8 ´ 50 mL
J-TB-056
24 ´ 15 mL J-TB-057