Appendix C: Diagnostics; Overview; Diagnostics/User Messages Chart - Beckman Coulter Avanti JXN-26 User Manual

Avanti jxn series high performance centrifuges for in vitro diagnostic use
Hide thumbs Also See for Avanti JXN-26:
Table of Contents

Advertisement

Overview

This section lists possible malfunctions and corrective actions. Maintenance procedures are
described in
CHAPTER 7, Maintenance and
For any maintenance not covered in this manual, contact your Beckman Coulter Representative for
assistance. USA customers can call 1-800-742-2345. For international contacts, see the website at
www.beckmancoulter.com or use the contact numbers on the inside front cover of this book.
NOTE
It is your responsibility to decontaminate the centrifuge, as well as any rotors and/or accessories,
before requesting service by your Beckman Coulter Representative.

Diagnostics/User Messages Chart

Refer to the chart below for a list of possible malfunctions and their corrective actions.
Table C.1 Diagnostics/User Messages Chart
Message
S1-S4/
Internal firmware error
S7
Internal firmware error
S9-10, S14/
Communications error
P5, S15/
Communications error
S108/
Network
communications error
PN B38322AA
Troubleshooting.
Definition/Result
Run coasts to a stop
Contact your Beckman Coulter Representative.
Run stops using maximum
Contact your Beckman Coulter Representative.
brake
Run coasts to a stop
Contact your Beckman Coulter Representative.
Run stops using maximum
Attempt to clear error. If problem persists,
brake
contact your Beckman Coulter Representative.
Connection to external
Make sure external devices are powered on.
devices lost
APPENDIX C
Diagnostics
Recommended Action
C-1