Entering Acceleration And Deceleration Times - Beckman Coulter Avanti J-HC Instructions For Use Manual

High-capacity centrifuge
Hide thumbs Also See for Avanti J-HC:
Table of Contents

Advertisement

Your selection (

Entering Acceleration and Deceleration Times

The centrifuge provides a choice of two acceleration rates (one for ARIES rotors) and four
deceleration rates (three for ARIES rotors) to protect the gradient and sample-to-gradient interface.
Table 2.1
describes these rates. If no rate is selected, the centrifuge automatically accelerates and
decelerates at maximum rate.
Table 2.1 Acceleration and Deceleration Settings
Setting
Maximum acceleration
Slow acceleration
(not available with
ARIES rotors)
Maximum deceleration
Slow deceleration
Timed deceleration
Off
(not available with
ARIES rotors)
Acceleration
1
Press
A/D
Accel:
The
2
Press the appropriate soft key. (
If you choose
PN J3HC-IM-5AD
YES
or
NO
) appears to the right of the
Rotor : JS-5.0
Speed : 5,000
RPM
=
Time :
00:00
Temp : 20°C
Max Temp : 24°C - YES
Accel : MAX
Decel : MAX
Overtemp shutdown active? YES or NO
Description
Full torque is used from 0 RPM until set speed is reached.
Approximately 2 minutes from 0 to 500 RPM. (Time depends on the mass of the
rotor in use; larger rotors take longer.) Above 500 RPM, full torque is used until
set speed is reached.
Full brake from set speed to near 0 RPM. Reduced brake is used during the last
few RPM until 0 is reached to minimize sample disturbance.
Reduced torque from set speed to 500 RPM, then approximately 2 minutes from
500 to 0 RPM. Deceleration to 500 RPM takes about twice as long as with
maximum deceleration. (The exact time depends on the mass of the rotor in
use; larger rotors take longer.)
Full brake from set speed to 500 RPM. From 500 to 0 RPM, deceleration can take
from 1 to 10 minutes (time is set by the user).
No brake is used. Rotor coasts from set speed to 0 RPM. Can take up to 1 hour,
depending on the mass of the rotor in use and the set speed.
.
MAX
SLOW
field flashes.
,
MAX
MAX
or
SLOW
, your selection appears in the
Max Temp:
YES
x G
NO
TIME
, and
soft keys appear.
is the only acceleration setting available with ARIES rotors.)
Accel:
Operation
Manual Operation
field.
field.
2
2-11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents