Operation
Action
1. Press [
].
TEMP
2. Press [ ] or [ ] until
the required temperature
appears, then press [

SETTING ACCELERATION AND DECELERATION RATES

Setting Acceleration Rate
Action
1. Press [
ACCEL
2. Press [ ] or [ ] until
the required acceleration
rate appears, then press
[
].
ENTER
3-6
Result
The last digit of the TEMP °C display flashes.
Entered temperature appears on the TEMP °C display.
].
ENTER
You can select one of ten acceleration rates and ten deceleration rates
to protect the gradient and sample-to-gradient interface. Select rates
depending on the type of run you are performing—for pelleting runs,
where sample mixing is not a concern, maximum deceleration can be
used. If running delicate gradients, a lower setting may be needed.
Typical acceleration and deceleration rates are listed in Table 3-1. If
no rate is selected, the centrifuge automatically accelerates and decel-
erates at maximum.
Result
].
The ACC/DEC display flashes.
Entered number appears on the ACC/DEC display.