Operation
Entering Run Parameters
Table 3.2 Deceleration Settings
Maximum Deceleration
Full brake is used from set
speed to near 0 rpm, to
bring the rotor to a stop in
the shortest possible time.
Reduced brake is used
during the last few rpm
until 0 is reached, to
minimize disturbance of the
sample.
Application: pelleting
Figure 3.1 Acceleration and Deceleration Settings Graph
1
500
0
1. Rotor Speed (RPM)
2. Time
3. Max Accel
4. Slow Accel
Setting the Acceleration Rate
NOTE
The default setting for both acceleration and deceleration is maximum (
lights are already lit, no input is required. To choose a different setting, follow the steps below.
3-8
Reduced torque is used from set speed to
500 rpm, causing deceleration to 500 rpm to
take twice as long as maximum deceleration.
(The exact rate depends on the mass of the
rotor in use. Larger rotors will take longer.)
Deceleration time from 500 to 0 rpm is
approximately 2 minutes (larger rotors take
longer).
Application: gradients, pelleting
5
3
4
2
5. Max Decel
6. Run Speed
7. Off (no brake)
8. Slow Decel
Slow Deceleration
7
8
Off
No brake is used. Rotor
coasts to 0 rpm. This
process can take up to
1 hour, depending on the
rotor in use and the speed
at the start of
deceleration.
Application: gradients
6
). If the
indicator
MAX
MAX
J326XP-IM-5AB