Fine Span Adjustment
4. FINE SPAN (FINE SPAN ADJUSTMENT)
If Fine Span Adjustment is desired, place a calibrated test weight on the scale and with the
SCALE 1 CALIBRATION MENU displayed, press the 4 key and then the ENTER key. The
display will change to show SCALE 1 FINE SPAN ADJUST. Note that ADJ = XXXXXXX
is the current displayed weight.
ADJ =
 Press UP (Navigation Keys  UP Arrow) to increase the span.
 Press DOWN (Navigation Keys  Down Arrow) to decrease the span.
 Press EXIT (Navigation Keys  Left Arrow) to save the new setting and return to the
SCALE 1 CALIBRATION MENU.
5. HI RES (HIGH RESOLUTION ADJUSTMENT)
If High Resolution Weight adjustment is desired, place a calibrated test weight on the scale
and with the SCALE 1 CALIBRATION MENU displayed, press the 5 key and then the
ENTER key. The display will change to show SCALE 1 HIRES ADJUSTMENT. Note that
HIRES=XXXXXX is the active weight in "high resolution" mode (in 1/10 interval).
ZERO :zero scale
HIRES=XXXXXX
 Press UP (Navigation Keys  UP Arrow) to increase the span.
 Press DOWN (Navigation Keys  Down Arrow) to decrease the span.
 Press EXIT (Navigation Keys  Left Arrow) to save the new setting and return to the
SCALE 1 CALIBRATION MENU.
 Press the ZERO key to zero the scale.
 Press the PRINT key to print the weight (followed by the text TEST).