Your 204 indicator has been thoroughly tested and calibrated before being shipped to you. If
you receive the indicator with a scale, calibration is not necessary. If the indicator is being
connected to a scale for the first time or recalibration is necessary for other reasons, proceed
as indicated.
Calibration of the 204 indicator is accomplished
entirely by the keypad. To enter the setup and
calibration mode:
1. With the power off, remove the Calibration
Access Screw on the upper left corner of the
rear panel, see Figure No.8.
2. With the screw removed, insert a small non-
metallic tool into the screw hole and press
and hold the calibration switch.
3. Press the ON/OFF key.
4. The display will show int=. The indicator is
now ready for setup and calibration.
During the setup and calibration process it will be necessary to enter operational
parameters via the 204's keyboard. Pressing the PRINT key will cause the data entered
or displayed to be retained and the 204 will advance to the next prompt. The cursor
location is identified by the blinking character and can be advanced to the left to the next
position by pressing the ASTERISK key. Pressing the UNITS key will change the
blinking character to the next value.
int= (Scale Interval)
With the display showing int= press the ASTERISK key to show the current setting. Press
the UNITS key until the proper scale interval (1, 2, 5, 10 or 20) is displayed and then press the
PRINT key to store the displayed value and proceed to the next prompt.
Flash= (Flash)
To display FLASH, press the calibration switch while the display is showing the prompt
int= (Scale Interval). With the display showing FLASh=, press the PRINT key. The
display will change to show 0 (0=NO). Press the PRINT key to proceed to the next prompt
Unit= (Weighing Units). To return to the int= prompt, start the setup and calibration
process over.
THE FLASH UPDATE OPTION IS AVAILABLE IN THE 204 INDICATOR AND IN
NCONTROL WITH REV 2.0 OR GREATER.
Unit= (Weighing Units)
With the display showing Unit= press the UNITS key to show the current setting. If the value
shown is acceptable, press the PRINT key again to save it, otherwise press the UNITS key to
enter the new weighing units and press the PRINT key to save the new setting.
0 = None
1 = Pounds Only
2 = Kilograms Only
8555-M313-O1 Rev H • 204 Installation & Technical
14

SETUP AND CALIBRATION

Figure No. 8
3 = Pounds/Kilograms
4 = Kilograms/Pounds
5 = Ounces Only
Calibration
Switch
6 = Grams Only
Calibration
Access
Screw