PRESET WEIGHT COMPARATORS SETUP
Mode of Operation = 3 (Preset Weight Comp)
With SETUP MENU #1 displayed the current setting for the 10. MODE OF OP= parameter
will be shown. Press the 1 and 0 keys and then the ENTER key. The display will change to
show the Mode Of Operation Menu. Press the 3 key and then the ENTER key to
select Preset Weight Comp. The display will change to show the Preset Weight
Comparator Menu below.
1. OUTPUTS=XX
2. BAL PRNT=XXX
Enter Selection: 0
1. OUTPUTS (NUMBER OF PRESET WEIGHT COMPARATORS "PWC")
With the Preset Weight Comparator Menu displayed the current setting for the
OUTPUTS= parameter will be shown. Note that XX is the current value. If the
setting displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter
the value for the number of PWC outputs and then press the ENTER key to save the
new setting and return to the Preset Weight Comparator Menu. Allowable values are
1 through 16.
The output state of each PWC before reaching the preset weight is set in the DIO SETUP
MENU. Refer to the DIO SETUP (PWC Status) section of the CALIBRATION AND SETUP
section of this manual.
The relay state is based on using Cardinal Scale relays (p/n 6850-1013). If using
relays other than those supplied by Cardinal Scale, refer to the relay manufacturers
specifications.
NOTE: When power to the 225 is lost, the output returns to a Low State condition.
94
Preset Weight Comparator Menu
8200-M698-O1 Rev B  225 Installation and Technical
^