Dual-Point without Zero (False Zero) Calibration
CAL 1= (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and CAL 1=0.0. This is the first of two calibration steps. This weight is
the TEST load (weight).
1. Place the weights on the scale.
2. Using the numeric keys, enter the value of the TEST load (weight) and then press the
ENTER key.
3. The display will flash CALIBRATING... for a few seconds and then change to
show CAL 2=.
CAL 2= (SECOND CALIBRATION WEIGHT)
The display will show CAL 2=XXXXXXX. This is the second of two calibration steps.
1. Remove the weights on the scale, and then press the NET/GROSS key.
2. The display will change to show the SCALE 1 CALIBRATION MENU. NOTE: If Span
Adjustment is required, proceed to the FINE SPAN ADJUSTMENT on page 70.
Single-Point for Span Only (Last Zero) Calibration
CAL 1= (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and CAL 1=0.0. This is the first of two calibration steps. This weight is
the TEST load (weight).
1. Zero the scale, and then place the weights on the scale.
2. Using the numeric keys, enter the value of the TEST load (weight) and then press the
ENTER key.
3. The display will flash CALIBRATING... for a few seconds and then change to
show CAL 2=.
CAL 2= (SECOND CALIBRATION WEIGHT)
The display will show CAL 2=XXXXXXX. This is the second of two calibration steps.
1. Remove the weights on the scale, and then press the NET/GROSS key.
2. The display will change to show the SCALE 1 CALIBRATION MENU. NOTE: If Span
Adjustment is required, proceed to the FINE SPAN ADJUSTMENT on page 70.
CALIBRATION AND SETUP, CONT.
8200-M698-O1 Rev B 225 Installation and Technical
81