Troubleshooting; Error Codes - Cardinal 225 Installation And Technical Manual

Table of Contents

Advertisement

Error Codes

The Model 225 indicator is equipped with software that indicates when an error in the
operation takes place. The following lists the error codes displayed by the 225 along with their
meaning. Should you encounter an error code, please refer to this list for the cause.
ANALOG HIGH
¤
¤
¤
¤
1. The load cell input is above the allowed range of the indicator.
CORRECTIVE ACTION: Check for improper load cell wiring, excessive load, and for
output of 1 to 40mV.
2. Load cell or circuit failure.
CORRECTIVE ACTION: Consult your scale service provider.
ANALOG LOW
¤
¤
¤
¤
1. The load cell input is below the allowed range of the indicator.
CORRECTIVE ACTION: Check for improper load cell wiring and for output of 1 to 40mV.
2. Load cell or circuit failure.
CORRECTIVE ACTION: Consult your scale service provider.
OVERCAP
¤
¤
¤
¤
The load on the scale exceeds the scale capacity (105% of capacity if USA=YES or plus 9
divisions if USA=NO). May indicate miscalibration.
CORRECTIVE ACTION: Remove the over capacity load from the scale platform. Consult
your scale service provider regarding recalibration.
CHECK MATL
In a Digital Fill Control or Batcher operation configured for decumulative operation
(ACCUMLATE=YES) there is not enough material to start the operation.
CORRECTIVE ACTION: Determine the reason for the error display and take the
appropriate corrective action.
CHECK ZERO
In a Digital Fill Control or Batcher operation with DUMP GATE=YES and AUTO
TARE=NO the weight is above the zero tolerance value.
CORRECTIVE ACTION: Determine the reason for the error display and take the
appropriate corrective action.

TROUBLESHOOTING

¤
¤
¤
¤
¤
¤
¤
¤
¤
8200-M698-O1 Rev B  225 Installation and Technical
125