Calibration Parameters

The calibration parameters (Swt, Sct and Zct) represent the test load (weight) and the
analog to digital readings at zero and test load. These values are established when the
scale is calibrated. To return the indicator to its present calibration settings (without using
test weights) enter the calibration parameters values.
1. Swt=XXXXXXX (SPAN WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed the current setting for the Swt=
parameter will be shown. Note that XXXXXXX is the current value for the calibrated test
load Span weight. If the setting displayed is acceptable, proceed to the next setup
parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter the
value for the calibrated test load Span weight and then press the ENTER key to save
the new setting and return to the SCALE 1 CALIBRATION MENU. Allowable values are
1 through 9,999,999. Note that the capacity cannot exceed 7 digits or 9,999,999.
2. Sct=XXXXXXX (SPAN COUNT)
With the SCALE 1 CALIBRATION MENU displayed the current setting for the Sct=
parameter will be shown. Note that XXXXXXX is the current value for the analog to
digital reading at test load Span count. If the setting displayed is acceptable, proceed to
the next setup parameter.
Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the test load Span count and then press the ENTER key to save the new
setting and return to the SCALE 1 CALIBRATION MENU. Allowable values are 1
through 9,999,999. Note that the capacity cannot exceed 7 digits or 9,999,999.
3. Zct=XXXXXXX (ZERO COUNT)
With the SCALE 1 CALIBRATION MENU displayed the current setting for the Zct=
parameter will be shown. Note that XXXXXXX is the current value for the analog to
digital reading at zero, the Zero count. If the setting displayed is acceptable, proceed to
the next setup parameter.
Otherwise, press the 3 key, the ENTER key and then using the numeric keys, enter the
value for the zero load Zero count and then press the ENTER key to save the new
setting and return to the SCALE 1 CALIBRATION MENU. Allowable values are 1
through 9,999,999. Note that the capacity cannot exceed 7 digits or 9,999,999.
84
CALIBRATION AND SETUP, CONT.