Your Model 210/215 indicator has been
thoroughly tested and calibrated before
being shipped to you. If you received the
indicator attached to a scale, calibration
is not necessary. If the indicator is being
connected to a scale for the first time or
recalibration is necessary for other
reasons, proceed as indicated.
The calibration switch is located on a
bracket on the inside of the enclosure
rear panel. You may gain access to this
switch simply by removing the calibration
switch access screw on the rear panel.
Refer to Figure No. 11.
During the setup and calibration process it is necessary to enter operational parameters via
the indicator's keypad. Pressing the ENTER key without entering a new value will retain the
current setting and advance to the next prompt. To change a setting, enter a new value and
press the ENTER key. This will save the new value and advance to the next prompt.
Pressing the ASTERISK key will "backup" to the previous prompt.
DO NOT operate the keypad with pointed objects (pencils, pens, etc).
Damage to keypad resulting from this practice is NOT covered under warranty.
Enter Setup Mode
To enter the setup mode, with the indicator ON, insert a small screwdriver or other tool through
the calibration switch access hole on the rear panel. Press and release the calibration switch.
The menu SetUP will be displayed. Continue to press and release the switch to rotate
through the beginning point for entering the setup mode.
SEtUP
Setup Mode (starts at USA prompt)
A-d
Analog to Digital Filtering (starts at dFLt= prompt)
CAL
Calibration (starts at CAL1 prompt)
Sio
Serial Input/Output (starts at bAUd prompt)
Print
Print Tab Settings (starts at POrt prompt)
F SPAn
Fine Span Adjustment
Hi rES
Display high-resolution weight mode
LoCoUt
Key lock out function
If you press the ENTER key at the SetUP prompt, you may proceed through to the next
section (up to and including f SPAn ) by pressing the ENTER key.
IMPORTANT! Setup may be interrupted at any time. ALL data previously
entered and finalized with the ENTER key will be retained in the non-volatile
memory.
8200-M411-O1 Rev D  210/215 Installation & Technical
20

SETUP AND CALIBRATION

Access
Screw
Figure No. 11