Your Model 200 Weight Indicator has been
thoroughly tested and calibrated before being
shipped to you. If you received the indicator
attached to a scale, calibration is not
necessary. If the indicator is being
connected to a scale for the first time or
recalibration is necessary for other reasons,
proceed as indicated.
The calibration switch is located on the
enclosure rear panel. You may gain access
to this switch by removing the 2 screws
securing the calibration switch access cover
on the rear panel. Refer to Figure No. 11.
During setup and calibration it will be necessary to enter operational parameters using the
indicator keypad.
Pressing the TARE/
retained and the indicator to advance to the next prompt.
The functions of numeric keys are replaced by using the UNITS/
the ASTERISK/
The cursor location is identified by the blinking character and can be advanced to
the next position (left) by pressing the UNITS/
Pressing the ASTERISK/
next value. Continue to press this key to toggle between the available values for the
setup parameter.
Pressing the ASTERISK/
value) is displayed, will "backup" to the previous prompt.
DO NOT operate the keypad with pointed objects (pencils, pens, etc).
Damage to keypad resulting from this practice is NOT covered under warranty.
Enter Setup Mode
To enter the setup mode, with the indicator ON, insert a small screwdriver or other tool through
the calibration switch access hole on the rear panel. Press and release the calibration switch.
The menu SetUP will be displayed. Continue to press and release the switch to rotate
through the beginning point for entering the setup mode.
SEtUP
A-d
CAL
SSt
Sio
Print
F SPAn
Hi rES
LoCoUt
option
If you press the TARE/
section (up to and including f SPAn ) by pressing the TARE/

SETUP AND CALIBRATION

key will cause the data entered or displayed to be
ENTER
keys.
UP ARROW
UP ARROW
UP ARROW
Setup Mode (starts at USA prompt)
Analog to Digital Filtering (starts at dFLt= prompt)
Calibration (starts at CAL1 prompt)
Setup Guardian Scale (starts at tLoU prompt)
Serial Input/Output (starts at Sio prompt)
Print Tab Settings (starts at POrt prompt)
Fine Span Adjustment
Display high-resolution weight mode
Key lock out function
Option Card Configuration (only displayed when option card is installed)
key at the SetUP prompt, you may proceed through to the next
ENTER
8200-M035-O1 Rev C  200 Installation & Technical
Calibration
Access Cover
Screws
LEFT ARROW
key will change the blinking character to the
key when a setup parameter (not a parameter
ENTER
Figure No. 11
LEFT ARROW
key.
key.
and
19